Keeping your world up and running.®

Fluke i6000sFlex AC Current Probe

The i6000s Flex is an ac current clamp utilizing the Rogowski principle. Used to measure currents up to 6000 A when used in conjunction with oscilloscopes, recorders or data loggers. Cable Accessories and Upgrades. Flexible current clamp.

Fluke i6000sFlex AC Current Probe
Overview
Specs
Models
Accessories
Resources
  • The i6000sflex is an AC current probe utilising the Rogowski principle. It can be used to measure currents up to 6000 A when used in conjunction with oscilloscopes, recorders or data loggers. The flexible and lightweight measuring head allows quick and easy installation in hard to reach areas.

  • Electrical Characteristics
    Current ranges 60 A / 600 A / 6000 A AC rms
    Output sensitivity (AC coupled) 50 mV /5 mV / 0.5 mV per A
    Load impedance 100 kOhm minimum
    Accuracy (at +25°C) ±1% of range (45 – 65 Hz)
    Linearity (10% to 100% of range) ±0.2% of reading
    Noise 8 mV AC rms (60 A)
    2 mV AC rms (600/6000 A)
    Frequency range 10 Hz to 50 kHz (-3 dB)
    Phase error < ±1° (45 – 65 Hz)
    ±10° (at 20 kHz)
    Position sensitivity (with cable > 25 mm from the coupling) ±2% of range
    External field (with cable > 200 mm from the head) ±1% of range
    Power supply 2 x AA MN 1500 LR6 alkaline
    400 Hours, low battery indicator or dedicated ext power supply
    Temperature coefficient ±0.08% of reading per ° C
    Working voltage (see Safety Standards section) 600 V AC rms or DC
    General Characteristics
    Head cable length (double insulated) 610 mm (24 in) I6000s Flex-24
    915 mm (36 in) I6000s Flex-36
    Head cable diameter 14.3 mm
    Head bend radius 38.1 mm
    Cable length (head to electronics) 2 m
    Output connection 0.5 m cable terminated with safety BNC connector
    Supplied with 4 mm safety plug adaptor
    Operating temperature range -20 to +90°C (head)
    -20 to +85°C (electronics)
    Storage temperature range -40 to +105°C (head)
    -20 to +85°C (electronics)
    Operating humidity 15% to 85% (non condensing)
    Weight 180 g (head), 190 g (electronics)
    Safety Standards
    BS EN 61010 1: 2001
    BS EN 61010 2-032: 2002
    BS EN 61010 031: 2002
    600 V rms, Category III, Pollution degree 2  
    EMC standards BS EN 61326: 1998 +A1, A2, and A3
    Use of the probe on uninsulated conductors is limited to 600 V AC rms or DC and frequencies below 1 kHz
    • Model
      Model Name Description  
      Fluke-i6000s Flex Fluke i6000sFlex AC Current Probe
      AC Current Probe
    • Manuals
      Title  
      i6000s Flex 使用说明单 Download
      i6000s Flex Instruction Sheet Download
      i6000s Flex Fiche d´instruction Download
      i6000s Flex Anleitungsblatt Download
      i6000s Flex Foglio di istruzioni Download
      i6000s Flex Folha de instruções Download
      i6000s Flex Hoja de instrucciones Download
      i6000s Flex Instruction Sheet Supplement Download
      i6000s Flex Statement of Memory Volatility Download