Fluke 9132/9133 Portable Infrared Calibrator

Our portable IR calibrators provide stable blackbody targets for calibrating noncontact IR thermometers from –30°C to 500°C

Get a Free Demo
See All Models »
Overview
Features
Reviews
Specs
Models
  • Precision when you need it for infrared temperature calibration

    • Certify IR pyrometers from –30°C to 500°C (–22°F to 932°F)
    • Large 57 mm (2.25 in) blackbody target
    • RTD reference well for contact temperature measurement
    • Small, compact design

    Whether you’re using in-line or handheld infrared pyrometers, you need good calibration standards to verify their accuracy. Our portable IR calibrators provide stable blackbody targets for calibrating noncontact IR thermometers from –30°C to 500°C.

    These units feature a temperature controlled measurement surface with a diameter of 2.25" (57 mm). The target temperature can be controlled in set-point increments of 0.1° from –30°C to 500°C and a well is located directly behind the blackbody surface for contact calibration.

  • 9132

    For IR calibrations above normal ambient, the 9132 provides a stable measurement surface up to 500°C (932°F). Short heating and cooling times mean you won't have to wait long to get your work done.

    9133

    If you're calibrating IR guns at cold temperatures, you'll love our new 9133. With solid-state cooling technology, this new IR calibrator reaches –30°C (22°F) in normal ambient conditions. With a conveniently located dry gas fitting on the front bezel, ice build up on the target can be avoided. At the upper end of its range, the 9133 provides stable temperatures to 160°C (320°F).

    With heating and cooling times of about 15 minutes from ambient to either extreme, the 9133 gets you to temperature quickly and performs when it gets there.

  • Specifications 9132 9133
    Temperature range 50°C to 500°C
    (122°F to 932°F)
    –30°C to 150°C at 23°C ambient
    (–22°F to 302°F at 73°F ambient)
    Accuracy ±0.5°C at 100°C
    (±0.9°F at 212°F)
    ±0.8°C at 500°C
    (±1.4°F at 932°F)
    ±0.4°C (±0.72°F)
    Stability ±0.1°C at 100°C
    (±0.18°F at 212°F)
    ±0.3°C at 500°C
    (±0.54°F at 932°F)
    ±0.1°C (±0.18°F)
    Target size 57 mm (2.25 in) 57 mm (2.25 in)
    Target emissivity 0.95 (±0.02 from 8 to 14 mm) 0.95 (±0.02 from 8 to 14 mm)
    Resolution 0.1° 0.1°
    Heating time 30 minutes (50°C to 500°C) 15 minutes (25°C to 150°C)
    Cooling time 30 minutes (500°C to 100°C) 15 minutes (25°C to –20°C)
    Computer interface RS-232 I/O included with 9930 Interface-it software
    Power 115 V AC (±10%), 3 A or
    230 V AC (±10%), 1.5 A,
    switchable, 50/60 Hz, 340 W
    115 V AC (±10%), 1.5 A, or
    230 V AC (±10%), 1.0 A,
    switchable, 50/60 Hz, 200 W
    Size (H x W x D) 102 x 152 x 178 mm
    (4 x 6 x 7 in )
    152 x 286 x 267 mm
    (6 x 11.25 x 10.5 in)
    Weight 1.8 kg (4 lb) 4.6 kg (10 lb)
    NIST-traceable contact calibration Data at 50°C, 100°C, 200°C, 250°C, 300°C, 400°C, and 500°C Data at –30°C, 0°C, 25°C, 75°C, 100°C, 125°C, and 150°C
    • Models
      Model Name Description  
      Fluke 9132 Fluke 9132 Portable Infrared Calibrator
      50°C to 500°C
      Fluke 9133 Fluke 9133 Portable Infrared Calibrator
      –30°C to 150°C at 23°C ambient