Keeping your world up and running.®

Fluke 725 Multifunction Process Calibrator

Fluke multifunction calibration tools designed to calibrate almost anything. These calibration tools source and measure almost all process parameters.

Calibrating a transmitter with a Fluke 725 Multifunction Process Calibrator
  • Calibrating a transmitter with a Fluke 725 Multifunction Process Calibrator
Overview
Features
Reviews
Specs
Models/Accessories
Resources
  • “Simply" Powerful!

    Take along the Fluke 725 Multifunction Process Calibrator, and you're equipped to test and calibrate almost any process parameter. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.

    From the moment you pick it up, the Fluke 725 is ready to perform. Its simple controls, without menus, make operation easy.

    Want more precision, pulse sourcing, totalizing and more? Consider the Fluke 726 Precision Multifunction Calibrator.

    Need an intrinsically safe rating in your calibrator? Consider the Fluke 725Ex Intrinsically Safe Multifunction Process Calibrator.

    Want documenting and HART communication built into your calibrator? Consider the Fluke 754 Documenting Process Calibrator with HART.

  • The Fluke 725 Multifunction Process Calibrator is a powerful, yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter.

    • Small, streamlined shape, rugged, reliable design stands up to field use
    • Easy to read measure/source screen lets you view input and output simultaneously
    • Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors and transmitters
    • Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
    • Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
    • Source mA with simultaneous pressure measurement to conduct valve and I/P tests
    • Support flow meter testing with frequency and CPM functions
    • Perform fast linearity tests with auto- step and auto-ramp features
    • Power transmitters during test using loop supply with simultaneous mA measurement
    • Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 10 ms
    • Store frequently used test setups for later use

    Want more precision, pulse sourcing, totalizing and more? Consider the Fluke 726 Precision Multifunction Calibrator.

    Need an intrinsically safe rating in your calibrator? Consider the Fluke 725Ex Intrinsically Safe Multifunction Process Calibrator.

    Want documenting and HART communication built into your calibrator? Consider the Fluke 754 Documenting Process Calibrator with HART.

  • Measurement Accuracy
    Voltage DC 30.000 V 0.02%+ 2 counts (upper display)
    30.000 V 0.02%+ 2 counts (lower display)
    100.00 mV 0.02%+ 2 counts
    -10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector)
    Current DC 24.000 mA 0.02%+ 2 counts
    Resistance 0.0 to 400.0 Ω 0.1Ω (4-wire), 0.15Ω (2- and 3-wire)
    401 to 1500 Ω 0.5Ω (4-wire), 1Ω (2- and 3-wire)
    1500 to 3200 Ω 1Ω (4-wire), 1.5Ω (2- and 3-wire)
    Frequency 2.0 to 1000.0 CPM 0.05 % + 1 count
    1.0 to 1100.0 Hz 0.05 % + 1 count
    1.00 to 10.00 kHz 0.05 % + 1 count
    Sensitivity 1 V peak-to-peak-minimum
    Pressure Accuracy From 0.025% of range using any of 29 pressure modules.
    (For detailed specifications refer to pressure modules in options and accessories)
    Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
    Source Accuracy
    Voltage DC 100.00 mV 0.02%+2 counts
    10.000 V 0.02%+2 counts
    -10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector)
    Current DC 24.000 mA (Source) 0.02%+ 2 counts
    24.000 mA (Simulate) 0.02%+ 2 counts
    Resistance 15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
    401 to 1500 Ω 0.5 Ω(excitation current 0.05 to 0.8 mA)
    1500 to 3200 Ω 1 Ω(excitation current 0.05 to 0.4 mA)
    Frequency 2.0 to 1000.0 CPM 0.05%
    1.0 to 1100.0 Hz 0.05%
    1.00 to 10.00 kHz 0.25%
    Waveform 5 V p-p squarewave, -0.1 V offset
    RTDs and Thermocouples
    Measure accuracy NI-120 0.2 °C
    PT-100 (385) 0.33 °C
    PT-100 (393) 0.3 °C
    PT-100 (JIS) 0.3 °C
    PT-200 (385) 0.2 °C
    PT-500 (385) 0.3 °C
    PT-1000 (385) 0.2 °C
    Resolution: 0.1 °C
    J: 0.7 °C
    K: 0.8 °C
    T: 0.8 °C
    E: 0.7 °C
    R: 1.8 °C
    S: 1.5 °C
    B: 1.4 °C
    L: 0.7 °C
    U: 0.75 °C
    N: 0.9 °C
    Resolution J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F
    XK 0.6°C
    BP 1.2°C
    Source accuracy NI-120 0.2 °C
    PT-100 (385) 0.33 °C
    PT-100 (393) 0.3 °C
    PT-100 (JIS) 0.3 °C
    PT-200 (385) 0.2 °C
    PT-500 (385) 0.3 °C
    PT-1000 (385) 0.2 °C
    Resolution: 0.1 °C
    Note: Accuracy stated for 4-wire measurement.
    J: 0.7 °C
    K: 0.8 °C
    T: 0.8 °C
    E: 0.7 °C
    R: 1.4 °C
    S: 1.5 °C
    B: 1.4 °C
    L: 0.7 °C
    U: 0.75 °C
    N: 0.9 °C
    Resolution J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C
    XK 0.6°C
    BP 1.2°C
    Specifications
    Ramp functions Source functions: Voltage, current, resistance, frequency, temperature
    Ramps Slow ramp, Fast ramp, 25% step-ramp
    Loop power function Voltage: 24 V
    Accuracy: 10%
    Maximum current: 22 mA, short circuit protected
    Step functions Source functions: Voltage, current, resistance, frequency, temperature
    Steps 25% of range, 100% of range
    Environmental Specifications
    Operating temperature -10 °C to +55 °C
    Storage temperature -20 °C to 71 °C
    Operating altitude 3000 m
    Safety Specifications
    Agency approvals EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
    Mechanical & General Specifications
    Size 130 x 236 x 61 mm
    Weight 0.65 kg
    Batteries 4 AA alkaline batteries
    Battery replacement Separate battery compartment, accessable without breaking calibration seal
    Side port connections Pressure module connector, also used for remote real-time programming
    • Manuals
      Title  
      725 Users Manual Download
      724/725/726 Calibration Manual Download
      724/725/726 Calibration Manual Supplement Download
      718,719,724,725, 726 Safety Sheet Download
      725 Statement of Memory Volatility Download
      725 Users Manual Download
      725 Brukerhåndbok Download
      725 用户手册 Download
      725 Brugsanvisning Download
      725 Handleiding Download
      725 Bedienungshandbuch Download
      725 Manuale per l’uso Download
      725 Manual do usuário Download
      725 Users Manual Download
      725 Käyttöohje Download
      725 Mode d'emploi Download
      725 Användarhandbok Download
      725 Manual de uso Download
      718,719,724,725, 726 Sicherheitsdatenblatt Download
      718,719,724,725, 726 Folha de Segurança Download
      718,719,724,725, 726 Turvaohjeet Download
      718,719,724,725, 726 Fiche de sécurité Download
      718,719,724,725, 726 Säkerhetsdatablad Download
      718,719,724,725, 726 Safety Sheet Download
      718,719,724,725, 726 Sikkerhetsdatablad Download
      718,719,724,725, 726 Safety Sheet Download
      718,719,724,725, 726 安全数据单 Download
      718,719,724,725, 726 Sikkerhedsblad Download
      718,719,724,725, 726 Veiligheidsinformatieblad Download
      718,719,724,725, 726 Foglio delle informazioni di sicurezza Download
      718,719,724,725, 726 Safety Sheet Download
      718,719,724,725, 726 Hoja de seguridad Download
      725 Users Manual Supplement Download
    • Brochures
      Title  
      Brochure: Fluke 725Ex Intrinsically Safe Multifunction Process Calibrator Download
      Fluke 725 Multifunction Calibrator Download
    • Application Notes
      Title  
      A brief look at ATEX and its impact in the US Download
    • Demos
      Title  
      Take the Fluke 725 for a test drive with the virtual demonstrator. Download
×