Keeping your world up and running.®

Fluke 80T-IR/E Extended Range Infrared Surface Temperature Probe

Fluke 80T-IR/E Extended Range Infrared Surface Temperature Probe
Overview
Features
Models
Resources
  • The 80T-IR/E is compatible with any DMM with 10 MOhms input impedance, standard diameter banana jacks, and mV dc capability. Ideal for maintenance applications where object to be measured is hazardous to touch, is moving, easily contaminated, or difficult to reach.

  • Specifications
    Temperature range 32 ºF to 1000 ºF
    Accuracy ± 3% of reading or ±3ºC (±5ºF), whichever is greater
    18ºC to 28ºC (64ºF to 82ºF) ambient operating temperature
    Spot size ratio 4:1
    Minimum spot size 0.1 inch diameter
    Output signal 1mV per ºF
    Safety rating Complies with IEC Publication 1010-1-1990 including Amendment 1
    CSA C22.2 No. 231, ANSI/ISA-S82.01 and .03 Safety Standards
    • Model
      Model Name Description  
      Fluke 80T IR/E Fluke 80T-IR/E Extended Range Infrared Surface Temperature Probe
      Extended Range Infrared Surface Temperature Probe
    • Manuals
      Title  
      80T-IR/E Calibration Information Download
      80T-IR/E Instruction Sheet Download